08/28/2023

Pulse wave doppler Aliasing is common problem usually encountered while interrogating vessels with pulse wave.

Image (a) showed a typical example of pulse wave aliasing.

PRF was too low = 1395.

This caused wrapping of the signal. As the sampled velocity is above the Nyquist limit .

In image (b) Aliasing has been corrected by raising the PRF value to 6250. This is a major effective solution to pulse wave aliasing.

The sampled signal must be consistently below the Nyquist limit.

For more detailed explanation on pulse doppler techniques

👇 👇 👇 👇 👇

https://youtu.be/VHEjZnqSWfo